Z-1392-2018: Ingenuity TF PET/CT, Model No. 882442 Product Usage: The device is an integrated diagnostic X-ray Comput

Recall record Z-1392-2018 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1392-2018
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA record creation date
Apr 17, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Following a period of inactivity, the mass storage device may cause the acquisition console to become unresponsive. This may prevent an acquisition from proceeding.

Source facts

Action
On February 15, 2018, the recalling firm notified customers of the recall via letter. The notice stated that to avoid the hazard of performing a rescan or reinjection, all users must be informed not to view studies or access data on the Addonics MSD while a patient is being scanned. If a user is attempting to view information stored on the Addonics MSD and the console appears to be unresponsive, reset power on the Addonics MSD or pull out and reconnect its USB cable. It is recommended that the Addonics USB Mass Storage Device be turned off when not being accessed, and powered on only when data transfer is needed. Following data transfer, the Addonics USB Mass Storage Device should be turned off. Customers were instructed to inform those who need to be aware within their organization or any organization wehre potentially affected devices have been transferred; transfer the notice to other affected organizations; and maintain the recall notice with their system Instructions For Use until the correction is made to the system. Customers were advised that Philips will be providing an updated Mass Storage Device that does not experience the same issue that will be provided at no additional cost to the customer. The Field Correction will be released in Q1 2018. Customers who need further information or support regarding the recall were directed to contact their local Philips representative. Customer Care Solutions Center: 1-800-722-9377.
Initiated
2018-02-15
Root cause
Under Investigation by firm
Terminated
2019-11-21
Distribution
Worldwide Distribution - US Nationwide in the states of: OH and OR, and Foreign distribution to Australia, China, Cuba, France, and South Africa.
Recall number
Z-1392-2018
Recall status
Terminated
CDRH record ID
163048
FDA record created
2018-04-17
Recall event number
79696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗