Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.
Source facts
Action
On May 31, 2022, the firm distributed Urgent Medical Device Product Removal letters to affected customers. Customers were instructed to immediately segregate and return the product to Boston Scientific.
Initiated
2022-05-31
Root cause
Software Manufacturing/Software Deployment
Distribution
Domestic distribution to AR
CA
IL
ME
MI
NC
TX
Foreign distribution to Australia, Canada, and France.
Recall number
Z-1392-2022
Recall status
Open, Classified
CDRH record ID
194455