Skip to content

Public FDA record

Z-1392-2023: FilmArray Gastrointestinal (GI) Panel for FILMARRAY Systems, REF: RFIT-ASY-0116, (30 pouches/kit)

Recall record Z-1392-2023 for BioFire Diagnostics, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1392-2023
Company in FDA record
BioFire Diagnostics, LLC
FDA classification date
Apr 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to manufacturing issue, Gastrointestinal (GI) Panel may result in false negative results.

Source facts

Action
On March 8, 2023, Biomerieux ( Parent company of BioFire Diagnostics, LLC) issued an "Urgent: Medical Device Correction" notification to affected consignees via E-Mail. In addition, to notifying affected consignees about the recalled device, Biomerieux asked consignees to take the following actions: 1. Immediately examine your inventory for product identified in this recall (refer to Table 1). 2. Discontinue use and discard any remaining product in your possession that is subject to this recall. bioMerieux will replace the product at no charge in accordance with BIOFIRE s standard limited warranty. 3. Please complete the accompanying Acknowledgment of Receipt Form and confirm the number of pouches scrapped (if any). Return the Acknowledgment of Receipt to bioMerieux so that bioMerieux may acknowledge your receipt of this notification. 4. If you have further distributed this product, please identify any recipients, and notify them at once of this product recall. 5. For product replacement, as well as any other questions or concerns, please contact our customer support department at biofiresupport@biomerieux.com or via telephone by dialing +1.800.736.6354 and selecting option 5 for Product Technical Support. 6. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. " Complete and submit the report online. " Regular Mail or Fax: Download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178.
Initiated
2023-03-08
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AR, CA, DC, IL, KS, LA, MN, NC, NY, OK, WI.
Recall number
Z-1392-2023
Recall status
Open, Classified
CDRH record ID
199116
FDA classified
2023-04-14
Recall event number
91892

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗