Z-1392-2024: X-Guide Handpiece Adaptor Sleeve 1, Model P010701

Recall record Z-1392-2024 for X-NAV Technologies, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1392-2024
Company in FDA record
X-NAV Technologies, LLC
FDA classification date
Mar 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Handpiece Adaptor may have a manufacturing defect causing incorrect geometry. The incorrect geometry in turn causes the Handpiece Adaptor to not fit onto the doctor's dental handpiece.

Source facts

Action
An URGENT MEDICAL DEVICE RECALL dated 3/4/24 was sent to customers. Action To Be Taken by the User Identify if you possess any of the affected devices. The lot numbers of the affected devices are the following: 2308010012 2306020012 2310010012 If the lot number on your device is not one of the above, then that Adaptor Sleeve is not part of this Recall. If you have a defective device, quarantine the device so that it is prevented from being used. Finally, destroy the device or return to X-Nav Technologies, LLC. Return the response form. X-Nav will ship all affected users a new Handpiece Adaptor Sleeve to replace their defective unit free of charge. This notice needs to be passed on all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. (As appropriate) Please transfer this notice to other organizations on which this action has an impact. (As appropriate) Please maintain awareness on this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Please report all device-related incidents to the manufacturer, distributor or local representative, and the national Competent Authority if appropriate, as this provides important feedback.
Initiated
2024-03-04
Root cause
Use error
Distribution
Worldwide distribution - US Nationwide and the countries of China, Europe, Japan.
Recall number
Z-1392-2024
Recall status
Open, Classified
CDRH record ID
206520
FDA classified
2024-03-29
Recall event number
94175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗