Skip to content

Public FDA record

Z-1392-2025: Origin Data Management software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.1

Recall record Z-1392-2025 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1392-2025
Company in FDA record
Brainlab AG
FDA classification date
Mar 18, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Under certain circumstances, it can occur that patient records from unrelated individuals are automatically merged into one patient record without user notification.

Source facts

Action
An URGENT FIELD SAFETY NOTICE / PRODUCT NOTIFICATION letter was sent to customers beginning 2/10/25. User Corrective Action: According to our records, at least one of the affected software versions is installed at your site. Therefore, when handling patient data intended to be used with Brainlab Software, adhere to the following instructions: Sequence 1 (Gender) " Do not edit only the Gender field of any already merged/edited patient record. (In the Patient Selection application, such a patient record is displayed with a Restore button next to it.) " For patient records that have not already been merged/edited, adapt all necessary fields including Gender in one single operation using Edit. Only then can you merge additional patient records to the edited patient record, if applicable. (In the Patient Selection application, such a patient record does not have a Restore button next to it.) Sequence 2 (ID) " Do not edit a Patient ID to a value that was the original ID of a previously merged/edited unrelated patient record. To verify whether a Patient ID has been previously used, perform the following steps before making any edits that affect Patient ID: 1. Search for the Patient ID that you want to use as the target ID in the search field in Patient Selection (do not use the Advanced Search). 2. If entries with the Restore button are shown in the patient list, - AND the searched Patient ID is not displayed - AND the Patient Name does NOT correspond to the individual intended to be treated, do not proceed with editing or merging and contact Brainlab support. For merges/edits originating externally from a Hospital Information System (HIS) that is connected to the above-listed versions of Brainlab ODM software: " Contact your HIS administrator (for the ADT interface) and clarify whether ADT A40 (merge patient) messages are used to edit Patient IDs in the system. If Patient IDs are changed via ADT A40 messages, contact Brainlab support so that we can verify whether our s
Initiated
2025-02-10
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-1392-2025
Recall status
Open, Classified
CDRH record ID
212401
FDA classified
2025-03-18
Recall event number
96241

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗