Some data alarm flags do not transfer from the analyzer to the laboratory information system, since not all data flags are recognized by the instrument manager/middleware.
Source facts
Action
Consignees were notified by "Urgent Medical Device Correction" letter dated April 1, 2009 and instructed to add error codes to the Instrument Manager driver and to contact the manufacturer of their LIS for instructions on adding data alarm codes to the LIS. Consignees were also instructed to complete the attached fax form after adding the error codes to their system and fax to 1-888-345-4690. Contact Roche Diagnostics Technical Support at 1-800-428-2336 for questions about information contained in the notification letter.
Initiated
2009-04-01
Root cause
Component design/selection
Terminated
2009-12-07
Distribution
Nationwide Distribution.
Recall number
Z-1393-2009
Recall status
Terminated
CDRH record ID
80699
FDA classified
2009-05-20
Recall event number
51561