Open original FDA record ↗Recall record
- FDA identifier
- Z-1393-2011
- Company in FDA record
- Advanced Sterilization Products
- FDA classification date
- Feb 22, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The recall was initiated after Advanced Sterilization Products (ASP) discovered a non-conforming lot. The misaligned perforations may have caused the seal to be incomplete on some of the pouches, creating a potential breach for pouches being used to store sterilized equipment on shelves.
Source facts
- Action
- Advanced Sterilization Products (ASP) released Urgent: Product Recall notifications to its consignees identifying the affected product and explaining the reason for recall. The letter also discussed the impact of using this affected product and return instructions. The firm asked customers to examine their inventory, discontinue use of Tyvek Self-Seal Pouch with STERRAD Chemical Indicator 12332 and to return all unused affected products to Stericycle. Customers are to complete the Business Reply Card and return it to Stericycle.
- Initiated
- 2010-06-23
- Root cause
- Process control
- Terminated
- 2012-03-14
- Distribution
- Worldwide Distribution -- USA, Argentina, Australia, Canada, Chile, Czech Republic, Hong Kong, India, Japan, Panama, Peru and Puerto Rico.
- Recall number
- Z-1393-2011
- Recall status
- Terminated
- CDRH record ID
- 93317
- FDA classified
- 2011-02-22
- Recall event number
- 56367