Skip to content

Public FDA record

Z-1393-2022: NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035

Recall record Z-1393-2022 for Synapse Biomedical Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1393-2022
Company in FDA record
Synapse Biomedical Inc
FDA record creation date
Jul 18, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).

Source facts

Action
On May 31, 2022, the firm began to contact all affected customer sites. Customers were informed that the way the temperature storage conditions are stated on the labeling of these devices could lead to improper storage that, under certain circumstances, could damage the devices. Synapse will visit customer facilities to place a label with proper storage conditions over the current storage conditions on the device label. Customers who would rather not have a site visit may destroy their devices and Synapse will replace them. The firm will also provide a revised Surgeon Instruction Manual that includes a description of the storage temperature symbol.
Initiated
2022-05-31
Root cause
Labeling design
Terminated
2024-04-11
Distribution
Domestic distribution to AL CA CO FL GA IL MA MI MN NC NJ NY OH OK OR PA SC TX WV. Foreign distribution pending.
Recall number
Z-1393-2022
Recall status
Terminated
CDRH record ID
194278
FDA record created
2022-07-18
Recall event number
90386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗