Open original FDA record ↗Recall record
- FDA identifier
- Z-1393-2022
- Company in FDA record
- Synapse Biomedical Inc
- FDA record creation date
- Jul 18, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).
Source facts
- Action
- On May 31, 2022, the firm began to contact all affected customer sites. Customers were informed that the way the temperature storage conditions are stated on the labeling of these devices could lead to improper storage that, under certain circumstances, could damage the devices. Synapse will visit customer facilities to place a label with proper storage conditions over the current storage conditions on the device label. Customers who would rather not have a site visit may destroy their devices and Synapse will replace them. The firm will also provide a revised Surgeon Instruction Manual that includes a description of the storage temperature symbol.
- Initiated
- 2022-05-31
- Root cause
- Labeling design
- Terminated
- 2024-04-11
- Distribution
- Domestic distribution to AL CA CO FL GA IL MA MI MN NC NJ NY OH OK OR PA SC TX WV. Foreign distribution pending.
- Recall number
- Z-1393-2022
- Recall status
- Terminated
- CDRH record ID
- 194278
- FDA record created
- 2022-07-18
- Recall event number
- 90386