Open original FDA record ↗Recall record
- FDA identifier
- Z-1394-04
- Company in FDA record
- General Electric Med Systems
- FDA classification date
- Aug 24, 2004
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector's radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become loose, fail, or back out.
Source facts
- Action
- The detector head of a FX-80 Nuclear camera may fall from the gantry. Three potential causes have been determined. (1) A ball nut in the detector''s radial drive assembly can back out of the bearing block that contains it. (2) Misalignment of the radial drive assembly. (3) Mounting bolts on the radial drive assembly may become lose, fail, or back out.
- Initiated
- 2004-07-30
- Root cause
- Other
- Terminated
- 2006-07-24
- Distribution
- Inside the US. Manufactured and distributed from 1993 - 2000.
- Recall number
- Z-1394-04
- Recall status
- Terminated
- CDRH record ID
- 34485
- FDA classified
- 2004-08-24
- Recall event number
- 29756