Z-1394-05: Baxter System 1000 Single Patient Hemodialysis Systems, product codes SYS1000L3, SYS1000L3P, SYS1000L3T,

Recall record Z-1394-05 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1394-05
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Aug 24, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Operator's Manuals addenda are being issued to provide correct information concerning the correct syringe size to use for heparin infusion, the use of chemical disinfectants in the System 1000, TMP Advise, and Bicarbonate Value calculations.

Source facts

Action
Baxter sent Important Product Information letters dated 6/14/05 to all System 1000 and Arena Hemodialysis customers, to the attention of the Hemodialysis Administrator, via first class mail. Enclosed with the letter were copies of addenda to the Operator''s Manual for both the System 1000 and Arena instruments. The letter informed the accounts of the following problems: a) Improper Syringe Size - it is important to use the correct syringe size for heparin infusion to avoid improper infusion of heparin. The warning 'Ensure the instrument is properly configured for the type and size of heparin pump syringe being installed.' was added to the manual under the Priming the Dialyzer and Blood Lines section of Operation. b) Disinfection Infusion Monitor (System 1000 only) - Procedural steps have been added to the System 1000 manual to confirm the level in the disinfectant bottle prior to and after chemical disinfection to ensure that disinfection fluid has been utilized during the disinfection cycle. c) TMP Advise - the 'TMP Alarm in Dialyzer Mode' section in both the System 1000 and Arena manuals has been removed. d) Bicarbonate Value - the Dialysate Preparation section of the manual was revised to clarify the bicarbonate display of the System 1000 and Arena instruments. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898.
Initiated
2005-06-14
Root cause
Other
Terminated
2008-02-26
Distribution
Nationwide and internationally to Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, India, Indonesia, Korea, Mexico, Peru, Philippines, Singapore, Spain, Switzerland, Taiwan, Thailand and Turkey.
Recall number
Z-1394-05
Recall status
Terminated
CDRH record ID
39626
FDA classified
2005-08-24
Recall event number
32322

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗