The table base rotational brake that holds the table top may not always be engaged when any force or side impact is applied to the table top.
Source facts
Action
Toshiba sent a Urgent Medical Device Correction letter dated March 11, 2015, to all customers who purchased the CAT-880B catherization table for the Tosiba Infinix X-Ray Interventional System. The letter informed the customers of the problems identified and the actions to be taken. Customers were instructed to complete and return the attached form and fax it to (877) 349-3054 or email to raffairs@tams.com. Customers with questions were instructed to call (800) 421-1968 or contact their local Toshiba Representative at (800) 521-1968.
For questions regarding this recall call 714-730-5000.
Initiated
2015-03-11
Root cause
Device Design
Terminated
2015-09-18
Distribution
Nationwide Distribution
Recall number
Z-1394-2015
Recall status
Terminated
CDRH record ID
134598
FDA classified
2015-04-06
Recall event number
70748