Skip to content

Public FDA record

Z-1395-2011: SP Brand Strep A Cassette; Part Number: B1077-26; Lot #: STA0010037 The SP Brand Strep A Rapid Test Devi

Recall record Z-1395-2011 for Biosite Inc Dba Innovacon Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1395-2011
Company in FDA record
Biosite Inc Dba Innovacon Inc.
FDA classification date
Feb 22, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated after the Investigation of customer-reported complaints, testing of market retention and returned product, It was determined that this lot of product has demonstrated an elevated Incidence of non-specific binding, potentially causing false positive results. A false positive test result could lead to the administration of antibiotics that are unnecessary. This exposes the

Source facts

Action
The firm, Alere San Diego, sent a "URGENT MEDICAL DEVICE RECALL" letter dated January 17, 2011. The letter described the product, problem and action to be taken by customers. The customers were instructed to: (1) Immediately discontinue use of the affected lot (if applicable) and inventory remaining full or partial kits at their site (if applicable); (2) Destroy any remaining inventory of this lot and document destruction per their site requirements (if applicable); (3) Complete and FAX the enclosed Verification Form within 10 days to confirm their receipt of this notice and to receive replacement of their remaining inventory of this lot (if applicable). The customers were ask to please return the form even if they have no remaining inventory or do not have any involved product in order to confirm receipt of this notification; and (4) to please ensure that all users of the device have received a copy of this Urgent Medical Device Recall. Customers with questions about the information contained in the notice were instructed to contact Technical Services at 888-246.7483, option 2 or by e-mail attechservices@alere.com.
Initiated
2011-01-12
Root cause
Pending
Terminated
2012-06-26
Distribution
Nationwide distribution
Recall number
Z-1395-2011
Recall status
Terminated
CDRH record ID
97073
FDA classified
2011-02-22
Recall event number
57715

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.