Skip to content

Public FDA record

Z-1395-2015: GE Discovery MR950 MRI system

Recall record Z-1395-2015 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1395-2015
Company in FDA record
GE Healthcare
FDA classification date
May 1, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The alignment lasers are missing the labels required by radiation safety regulations.

Source facts

Action
GE Planned action(s) to repair defect or to bring product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 5 units consists of the following: 1. New labels will be mailed to all purchasers, with instructions to apply the label in the appropriate location on the unit, or a GE field engineer will be sent to each user site to complete the correction. 2. All mailings and service calls will be made at no cost to the purchaser. 3. The corrections will be completed by August 1, 2015. If you have any questions, please contact your local Service Representative or call 1 (800) 437-1171.
Initiated
2014-10-24
Root cause
Error in labeling
Terminated
2015-05-06
Distribution
US Distribution to the state of CA., and Internationally to Italy and Japan.
Recall number
Z-1395-2015
Recall status
Terminated
CDRH record ID
134519
FDA classified
2015-05-01
Recall event number
70701

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗