Open original FDA record ↗Recall record
- FDA identifier
- Z-1395-2015
- Company in FDA record
- GE Healthcare
- FDA classification date
- May 1, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The alignment lasers are missing the labels required by radiation safety regulations.
Source facts
- Action
- GE Planned action(s) to repair defect or to bring product into compliance: The Center for Devices and Radiological Health (CDRH) understands that the CAP involving 5 units consists of the following: 1. New labels will be mailed to all purchasers, with instructions to apply the label in the appropriate location on the unit, or a GE field engineer will be sent to each user site to complete the correction. 2. All mailings and service calls will be made at no cost to the purchaser. 3. The corrections will be completed by August 1, 2015. If you have any questions, please contact your local Service Representative or call 1 (800) 437-1171.
- Initiated
- 2014-10-24
- Root cause
- Error in labeling
- Terminated
- 2015-05-06
- Distribution
- US Distribution to the state of CA., and Internationally to Italy and Japan.
- Recall number
- Z-1395-2015
- Recall status
- Terminated
- CDRH record ID
- 134519
- FDA classified
- 2015-05-01
- Recall event number
- 70701