- Action
- The firm initiated their recall by letter on March 16, 2018. The notice stated:
"Under certain conditions, when using DICOM Worklist along with DICOM MPPS, a report for a bone density exam may be sent to PACS with the incorrect patient information in the DICOM header. The correct patient information will be listed on the DICOM report image; however, the report may appear under a different patients name in the PACS. There have been no injuries reported as a result of this issue.
You can continue using your system by disabling the DICOM MPPS feature as follows:
1. In enCORE, select Tools | Connectivity Options.
2. Select DICOM tab.
3. Press Configure Connections in DICOM Worklist section.
4. In the MPPS SCP Settings, clear the AE Title and the TCP/IP Address fields
5. Press OK to all.
A GE Healthcare representative will contact you to arrange for the correction."
For further questions, please call (608) 826-7492
- Initiated
- 2018-03-16
- Root cause
- Software design
- Terminated
- 2020-12-14
- Distribution
- Worldwide Distribution - USA (nationwide) including Puerto Rico and to the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Egypt, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Lebanon, Malaysia, Malta, Mexico, Morocco,
Norway, Poland, Portugal , Romania, Russia, Saudi Arabia, Spain, Switzerland, Taiwan, Tunisia, Turkey, United Arab, United Kingdom and Venezuela.
- Recall number
- Z-1395-2018
- Recall status
- Terminated
- CDRH record ID
- 163129
- FDA record created
- 2018-04-17
- Recall event number
- 79710