Instrument Kits for EPS plates may contain drill guides with incorrect drill depth range of 30mm - 60mm; not the correct drill depth range of 10mm - 40mm.
Source facts
Action
On 4/4/23, recall notices were mailed to customers and distributors who were asked to do the following:
1) Quarantine affected kits so they are ready for removal by the firm s sales reps or other contact.
2) If devices were transferred or further distributed, notify customers, or provide a customer list, so the recalling firm can notify these customers directly.
3) Complete and return the response form to msimpson@gramercyortho.com
Customers with questions can contact the firm at 855-436-2278, 9AM 5PM, Monday - Friday.
Initiated
2023-04-04
Root cause
Mixed-up of materials/components
Distribution
US Nationwide distribution in the states of CT, OR, NJ, AZ, CT, OH, AZ.
Recall number
Z-1395-2023
Recall status
Open, Classified
CDRH record ID
199491
FDA classified
2023-04-17
Recall event number
92051