Open original FDA record ↗Recall record
- FDA identifier
- Z-1396-2011
- Company in FDA record
- Hill-Rom, Inc.
- FDA classification date
- Feb 23, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Slings were improperly assembled and the loops that are used to support the sling when hung from a lift were too long, potentially putting the patient being lifted at risk of falling out.
Source facts
- Action
- The firm sent an URGENT MEDICAL DEVICE RECALL notice dated 02/09/2011 to Facility Risk Manager/Facility Administrators. The letter Identified the affected product, explained the recall situation, and asked firms to check their facility for the affected serial numbers. Consignees were asked to return the recalled products, and a replacement sling would be sent to them. Customers can contact Hill-Rom Technical Support at 800-445-3720, if they have any questions concerning this matter.
- Initiated
- 2011-02-09
- Root cause
- Pending
- Terminated
- 2011-10-07
- Distribution
- Worldwide Distribution -- US, Switzerland, Germany, Denmark, Finland, France, UK, Italy, Netherlands, Sweden, and Canada.
- Recall number
- Z-1396-2011
- Recall status
- Terminated
- CDRH record ID
- 97603
- FDA classified
- 2011-02-23
- Recall event number
- 57861