Z-1396-2025: CVAC Aspiration System, REF: CVC127020-1, and User Manual L00018. Intended for endoscopic examination/tre

Recall record Z-1396-2025 for Calyxo, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1396-2025
Company in FDA record
Calyxo, Inc.
FDA classification date
Mar 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Aspiration system, for endoscopic examination/treatment of urinary tract and kidney interior, to have labeling update adding additional instructions, for patients with high viscosity fluid in the kidney, not to continue providing fluid inflow in the presence of unresolved slow or absent fluid outflow because this can create intrarenal pressure imbalance, and lead to excessive intrarenal pressure.

Source facts

Action
On 2/19/2025, correction notices were mailed to distributors and healthcare providers who were asked to do the following: 1) Notify all urologists utilizing the affected aspiration system in your institution that additional instructions are required. Please provide each of them with a copy of the additional instructions. 2) As another safety step, a firm representative will be contacting all users from your institution directly to offer training on these additional instructions. If you prefer to coordinate the training, please specifically state that preference in the acknowledgement. 3) Please acknowledge receipt of the notice, additional instructions, and updated training instructions MA00100.A by emailing qualitycontrol@calyxoinc.com, or calling 833-214-3354 and providing your name, title, institution and address.
Initiated
2025-02-19
Root cause
Device Design
Distribution
US Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV.
Recall number
Z-1396-2025
Recall status
Open, Classified
CDRH record ID
212589
FDA classified
2025-03-21
Recall event number
96340

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗