Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
Source facts
Action
Firm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.
Initiated
2005-07-25
Root cause
Other
Terminated
2007-01-31
Distribution
Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,
Recall number
Z-1397-05
Recall status
Terminated
CDRH record ID
40783
FDA classified
2005-08-24
Recall event number
32762