Z-1397-2023: Habib EndoHPB Bipolar Radiofrequency (RF) Catheter 8F (2.7mm). A radiofrequency (RF) catheter which provi

Recall record Z-1397-2023 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1397-2023
Company in FDA record
Boston Scientific Corporation
FDA classification date
Apr 18, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Cease use of the Habib EndoHPB Bipolar RF Catheter with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage. An inaccurate setting is currently provided in the HABIB EndoHPB Catheter's Instructions for Use ("IFU").

Source facts

Action
Boston Scientific issued Urgent Medical Device Field Safety Notice - Immediate Action Required Habib" EndoHPB Bipolar Radiofrequency Catheter IFU Generator Setting to US and Puerto Rico consignees sent via Federal Express Priority Overnight mail on March 2, 2023. Letter states reason for recall, health risk and provides important information about the use of the Habib EndoHPB Catheter when used with the ERBE VIO 3 Generator. Habib EndoHPB Catheter users should cease use with the ERBE VIO3 Generator immediately due to the potential for delivery of excessive energy leading to thermal injury or tissue damage as a result of the inaccurate setting currently provided in the HABIB EndoHPB Catheter s Instructions for Use ( IFU ). Boston Scientific recommends that patients who have been treated using the Habib EndoHPB Catheter with the ERBE VIO 3 Generator continue with their regular or routine check-ups and follow-up care. Please read this letter carefully and immediately post this information in a visible location near the product to ensure this information is easily accessible to all users of the device. Share this notice with any health professional within your organization that needs to be aware and to any organization where the potentially affected devices have been transferred. If you are a distributor, this notification must be forwarded to your customers to ensure notification is carried out to the end user level. If you are a facility that has sent products to another hospital or facility within your network, ensure that this notification is forwarded to them. If you have additional questions regarding this information or would like to report a clinical event, please contact your Boston Scientific representative
Initiated
2023-02-02
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Bulgaria, Canada, Costa Rica, Czech Republic, Egypt, Finland, France, Germany, Great Britain, Greece, Hong Kong, Hungary, Iceland, Iraq, Israel, Italy, Japan, Korea, Kuwait, Latvia, Lithuania, Mexico, Netherlands, New Zealand, Oman, Panama, Poland, Portugal, Romania, Russian Fed., Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Unit. Arab Emir.
Recall number
Z-1397-2023
Recall status
Open, Classified
CDRH record ID
198889
FDA classified
2023-04-18
Recall event number
91792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗