Skip to content

Public FDA record

Z-1398-06: GammaMed software program, version 5.07, vers. 5.08, vers.-usa, vers.-tst and vers.-row , part number GM1

Recall record Z-1398-06 for Varian Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1398-06
Company in FDA record
Varian Medical Systems
FDA classification date
Aug 10, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans manually that the user neglects to change the default step size and/or the origin or accidentally enters incorrect parameters.

Source facts

Action
The recalling firm notified healthcare practitioners using the medical device by letter on 4/17/06 and continuing through 5/1/06. The notification was flagged as a 'Medical Device Recall' and instructed users to adjust system parameters in the autoexec.bat file. The notification further recommended using the identified treatment plans and that manual input should be used for QA purposes only.
Initiated
2006-05-01
Root cause
Other
Terminated
2006-12-05
Distribution
Nationwide.
Recall number
Z-1398-06
Recall status
Terminated
CDRH record ID
46680
FDA classified
2006-08-10
Recall event number
35717

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗