Open original FDA record ↗Recall record
- FDA identifier
- Z-1398-2015
- Company in FDA record
- Siemens Healthcare Diagnostics, Inc.
- FDA classification date
- Apr 7, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
A positive bias beyond 5% resulting in a dilution check failure or an unnecessary dilution check correction if the positive bias is greater than 1% and less than 5%. The dilution check correction would be detected by QC; however, if not detected by QC, the calibration curve would be decreased and cause elevated sodium, potassium and chloride patient results.
Source facts
- Action
- An Urgent Medical Device Recall letter, dated February 2015, was provided to all Dimension system customers who received the affected lot to notify them that Siemens confirmed customer complaints of failing dilution checks when using QuikLYTE Dilution Check lot 4MD707. The letter also instructed them to immediately discontinue use and discard all inventory of Dimension QuikLYTE Dilution Check lot 4MD707. Siemens offered a no charge replacement with a non-impacted lot.
- Initiated
- 2015-02-24
- Root cause
- Under Investigation by firm
- Terminated
- 2017-02-01
- Distribution
- Worldwide Distribution.
- Recall number
- Z-1398-2015
- Recall status
- Terminated
- CDRH record ID
- 134171
- FDA classified
- 2015-04-07
- Recall event number
- 70627