Open original FDA record ↗Recall record
- FDA identifier
- Z-1398-2017
- Company in FDA record
- Toshiba American Medical Systems Inc
- FDA record creation date
- Mar 14, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
It was discovered that the generator of the system could possibly terminate the exposure prematurely during an examination
Source facts
- Action
- Toshiba America Medical Systems planned action to bring defect into compliance : 1. A notification letter will be sent to customers with affected systems. This letter includes the statement that the repairs will be made at no cost to the customer. 2. Modified firmware will be installed on the systems to prevent the defect from recurring. 3. The consignees will be contacted by a Toshiba service representative for an appointment when the modified firmware becomes available 4. The consignee will be informed that if this problem occurs before the new firmware is installed; they should reset the x-ray generator to complete the exam; and 5. The CAP will be completed by May 1, 2017. For further questions please call (714) 730-5000
- Initiated
- 2016-05-19
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2018-06-14
- Distribution
- USA ( nationwide ) Distribution to the states of : WA, OR, CA, OH, IL, CT, IA, KY, AL, LA, HI, PA, NJ, DE, MN, DC, NY, MI, SC, MD, TN, OK, WI, UT, ID, NV, TX, AL, IN, WV, KS, MO, CO, MT, MS, NC, GA, NH,UT, FL, AZ and ND.
- Recall number
- Z-1398-2017
- Recall status
- Terminated
- CDRH record ID
- 152450
- FDA record created
- 2017-03-14
- Recall event number
- 76175