Open original FDA record ↗Recall record
- FDA identifier
- Z-1399-06
- Company in FDA record
- Varian Medical Systems
- FDA classification date
- Aug 10, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Software control program for a medical device used in radiation treatment may cause practitioners to incorrectly administer the treatment plan to cancer patients. There is a hazard when entering the treatment plans manually that the user neglects to change the default step size and/or the origin or accidentally enters incorrect parameters.
Source facts
- Action
- The recalling firm notified healthcare practitioners using the medical device by letter on 4/17/06 and continuing through 5/1/06. The notification was flagged as a 'Medical Device Recall' and instructed users to adjust system parameters in the autoexec.bat file. The notification further recommended using the identified treatment plans and that manual input should be used for QA purposes only.
- Initiated
- 2006-05-01
- Root cause
- Other
- Terminated
- 2006-12-05
- Distribution
- Nationwide.
- Recall number
- Z-1399-06
- Recall status
- Terminated
- CDRH record ID
- 46880
- FDA classified
- 2006-08-10
- Recall event number
- 35717