Open original FDA record ↗Recall record
- FDA identifier
- Z-1399-2009
- Company in FDA record
- Siemens Medical Solutions Diagnostics
- FDA classification date
- May 26, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Design Flaw: The redesign of the Digital Fluidics PCB used in IMMULITE 2000 and IMMULITE 2500 is missing a jumper.
Source facts
- Action
- Siemens Healthcare Diagnostics Inc. issued an "Urgent: Device Recall" notice dated January 2009 mailed to all direct customers of the affected product. The letter described the problem, provided "work-around" instructions and also provided a Recall Response Form to be returned to the firm via fax. For further information, please contact your local Siemens Healthcare support representative or Siemens Healthcare Diagnostics Inc. directly at 1-973-927-2828.
- Initiated
- 2009-01-05
- Root cause
- Process design
- Terminated
- 2010-04-19
- Distribution
- Worldwide Distribution to states including MA, NY, PA, MD, GA, FL, OH, IN, MI, IL, MO, LA, UT, AZ and CA and countries of Canada, Czech Republic, Germany, Poland, Hungary, Belgium, Russia, Republic of South Africa, Greece, Israel, Turkey, Spain, Portugal, Denmark, France, Great Britain, India, China, South Korea and Mexico.
- Recall number
- Z-1399-2009
- Recall status
- Terminated
- CDRH record ID
- 77487
- FDA classified
- 2009-05-26
- Recall event number
- 50849