Z-1399-2019: AURIGA 30 GENERAL ZERO COST SYSTEM UPN: M0068S30GZ0 Product Usage: The Auriga XL 4007 Laser System an

Recall record Z-1399-2019 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1399-2019
Company in FDA record
Boston Scientific Corporation
FDA classification date
May 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Certain Auriga consoles may exhibit faster than normal power degradation due to a potential issue in the mechanical cooling system that could lead to a low power error (1301) being displayed on the screen. The console is designed to stop emitting laser pulses and enters a stand-by mode after a low power error message.

Source facts

Action
Boston Scientific issued to the user level the correction/removal via a customer notification letter via FedEx or hand delivery by a Boston Scientific agent on 4/9/19. Letter provides recommendations to the user until a field repair can be completed or a replacement console can be provided. A Boston Scientific representative will be in contact with you directly to schedule a field repair or replacement at your facility. The console can continue to be used during this time following the recommendations within the letter.
Initiated
2019-04-08
Root cause
Component design/selection
Terminated
2023-05-10
Distribution
Worldwide distribution to Australia, Belgium, Brazil, Chile, Colombia, Egypt, France, Germany, Great Britain, India, Indonesia, Iran, Italy, Lebanon, Malaysia, Mexico, Romania, Russia Fed., South Africa, South Korea, Spain, Taiwan, United Arab Emirates and Vietnam
Recall number
Z-1399-2019
Recall status
Terminated
CDRH record ID
172404
FDA classified
2019-05-22
Recall event number
82691

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗