Skip to content

Public FDA record

Z-1399-2024: BD Phoenix Yeast ID Panel is a self-inoculating (when used with a BD Phoenix Automated Microbiology Syste

Recall record Z-1399-2024 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1399-2024
Company in FDA record
Becton Dickinson & Co.
FDA classification date
Mar 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Esculin, a substrate used in other panels and normally in A16 well, was found in A17 well. A17 in yeast panels is supposed to be an empty well and is not a part of the normal ID evaluation. However, A17 is checked for fluorescence and if anything is found, the system will abort the panel with the message Fluorescent interference has been detected. Esculin is expected to trigger this fluorescence and test cancellation.

Source facts

Action
BD issued URGENT: Medical Device Product Correction Product Removal letter on 2/29/24 via courier service via tracked letter and/or e-mail. Letter states reason for recall, health risk and action to take: 1. Immediately inspect your inventory for the specific catalog and lot number listed above. Destroy affected product subject to the recall following your institution s process of destruction. 2. Ensure the contents of this notification are read and understood. 3. Share and post this customer letter within your facility network and forward to any customers to whom you may have distributed the product to ensure awareness. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements. BD will issue product replacements following receipt of the completed Customer Response Form. Contact: North American Regional Complaint Center Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com Product Advisory, Recall Questions, Product Complaints, Technical Questions
Initiated
2024-02-29
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of AR, BE, CO, IN, JP, PE, PY, SG, TW, UY.
Recall number
Z-1399-2024
Recall status
Open, Classified
CDRH record ID
206527
FDA classified
2024-03-29
Recall event number
94180

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗