Reason stated in the source
Certain TeleSentry Devices contain a Bad SD Card. When the SD Card goes bad, the TeleSentry Device cannot read or write data to the SD Card. The TeleSentry Device then stops functioning.
Source facts
- Action
- ScottCare sent an "URGENT MEDICAL DEVICE RECALL" letter dated January 19, 2012 to all affected customers. The letter identifies the product, problem, and the actions to be taken by the customers. The letter instructs customers to immediately discontinue use and quarantine any affected products. A Recall Acknowledgement Form was enclosed for customers to complete and return via fax to (216) 257-6129. Contact the firm at (216) 362-0550 ext. 113 for questions regarding this recall.
- Initiated
- 2012-01-19
- Root cause
- Component design/selection
- Terminated
- 2013-09-27
- Distribution
- Worldwide Distribution-USA (nationwide) including the states of CA, DE, FL, MI, NJ, NV, NY, OH, PA and TX, and the country of Canada.
- Recall number
- Z-1400-2012
- Recall status
- Terminated
- CDRH record ID
- 107929
- FDA classified
- 2012-04-13
- Recall event number
- 61370