Z-1400-2019: Thio Med w/o Ind, w/Dex (9ml) 100/PK Product Usage: Recommended for use in qualitative procedures as a g

Recall record Z-1400-2019 for Remel Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1400-2019
Company in FDA record
Remel Inc
FDA record creation date
May 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for products to fail performance testing per the Instructions for Use (IFU) starting at 106 days after date of manufacturing. Expiration date is shortened from 9 months to 105 days.

Source facts

Action
Thermo Fisher Scientific/ Remel notified consignees on about 04/17/2019 via " URGENT MEDICAL DEVICE FIELD ACTION RECALL" letter. Instructions included to locate all affected in inventory, destroy any affected product identified, contact Customer Services or local distributor to arrange for replacements, notify all customers if product was further distributed, and complete and return the Field Action Response form. Customers were also instructed that the requirement for review of reported test results using product beyond the updated expiry date should be determined by the appropriate technical expert at each customer location. Thermo Fisher Scientific/Remel expanded the recall and notified consignees on about 05/08/2019 via "URGENT MEDICAL DEVICE FIELD ACTION RECALL AMENDED NOTICE: ADDITIONAL IMPACTED LOTS" letter. The letter contained the same instructions as the initial letter, but the updated letter provided the additional impacted lots and requested that consignees respond again confirming receipt and understanding of the expansion. Questions or concerns can be directed to Technical Services Department at 800-255-6730 (US) or 913-888-0939 (International).
Initiated
2019-04-17
Root cause
Under Investigation by firm
Terminated
2020-07-10
Distribution
Worldwide Distribution -US Nationwide to AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, Puerto Rico. International distribution to Canada.
Recall number
Z-1400-2019
Recall status
Terminated
CDRH record ID
172346
FDA record created
2019-05-22
Recall event number
82683

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗