Mindray DS USA, Inc., dba Datascope Patient Monitoring
FDA classification date
May 27, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification
Reason stated in the source
Software Anomaly: Software anomaly can occur when the Panorama is attempting to play an audio file, either a physiological or technical alarm sound, at which time the system software may cause the system to freeze due to an interaction in the alarm system.
Source facts
Action
Urgent Product Field Correction letters, dated March 27, 2009, were sent by certified mail, return requested, and signature required. The letters identified the affected product and the problem. It also stated the adverse effect on patients and that customers' Panorama software will be upgraded to a current software revision. A Datascope Service Representative will contact customers to arrange for the software upgrade of the Panorama(s) in their facility. This will be done at no cost to the customer.
Initiated
2009-04-10
Root cause
Software design
Terminated
2010-08-03
Distribution
Worldwide Distribution -- USA and Australia.
Recall number
Z-1401-2009
Recall status
Terminated
CDRH record ID
81474
FDA classified
2009-05-27
Recall event number
51863