Open original FDA record ↗Recall record
- FDA identifier
- Z-1401-2013
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- May 25, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm initiated recall of Alere" hCG Combo Cassette (20/10 mIU/mL) kit PN 92215, lots hCG1110133 and hCG1110135, because an investigation has indicated a portion of these Alere" hCG Combo Cassette (20/10 mIU/mL) PN 92215 kit lots were incorrectly packaged with the Alere" hCG Cassette (20 mIU/mL) PN 92210 Product Insert. The Alere" hCG Cassette (20 mIU/mL) Product Insert contains information for
Source facts
- Action
- Alere initiated this recall by sending recall notification letters, titled "URGENT MEDICAL DEVICE RECALL", dated April 16, 2013 via via fax, email or direct mail. The letter informed customers of the product recalled, reason for recall, customer required action, and contact information, "Alere Technical Services at 866-216-0094 or by e-mail at lateral.flow.support@alere.com.". A verification form was attached to the notification letter.
- Initiated
- 2013-04-16
- Root cause
- Packaging
- Terminated
- 2013-10-18
- Distribution
- United States Nationwide Distribution
- Recall number
- Z-1401-2013
- Recall status
- Terminated
- CDRH record ID
- 117797
- FDA classified
- 2013-05-25
- Recall event number
- 65048