Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially
requiring surgical or medical/pharmacological intervention
Source facts
Action
Terumo issued URGENT Medical Device Safety Alert on 2/20/24 via 2-day FedEx delivery.
Letter states reason for recall, health risk and action to take:
Do not use the valve included in the kit and discard. The clinician could implement clinical changes such as:
1. Place a stopcock in a closed position or a non-vented cap on the luer lock.
2. Replace the line to the reservoir that has the impacted high pressure check valve in it with another equivalent over-pressure line. The High Crack Check Valve (PN 205307A) is not assembled directly into the finished good and can be replaced with a sterilized over-pressure line.
contact Terumo CVS with any questions or concerns:
Terumo CVS Customer Service: 1.800.521.2818
Monday Friday, 8 a.m. 6 p.m. ET
Terumo Recall Fax: 732-507-9046 or 732-507-9046@efax.com
Terumo Recall Email: tcvs.recall@terumomedical.com
Initiated
2024-02-20
Root cause
Process control
Distribution
worldwide distribution - US Nationwide and the countries of Canada, Chile, Colombia,
Mexico, South Korea, Taiwan, Thailand.
Recall number
Z-1401-2024
Recall status
Open, Classified
CDRH record ID
206264
FDA classified
2024-03-29
Recall event number
94083