Z-1401-2024: Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 6297

Recall record Z-1401-2024 for Terumo Cardiovascular Systems Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1401-2024
Company in FDA record
Terumo Cardiovascular Systems Corporation
FDA classification date
Mar 29, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

Source facts

Action
Terumo issued URGENT Medical Device Safety Alert on 2/20/24 via 2-day FedEx delivery. Letter states reason for recall, health risk and action to take: Do not use the valve included in the kit and discard. The clinician could implement clinical changes such as: 1. Place a stopcock in a closed position or a non-vented cap on the luer lock. 2. Replace the line to the reservoir that has the impacted high pressure check valve in it with another equivalent over-pressure line. The High Crack Check Valve (PN 205307A) is not assembled directly into the finished good and can be replaced with a sterilized over-pressure line. contact Terumo CVS with any questions or concerns: Terumo CVS Customer Service: 1.800.521.2818 Monday Friday, 8 a.m. 6 p.m. ET Terumo Recall Fax: 732-507-9046 or 732-507-9046@efax.com Terumo Recall Email: tcvs.recall@terumomedical.com
Initiated
2024-02-20
Root cause
Process control
Distribution
worldwide distribution - US Nationwide and the countries of Canada, Chile, Colombia, Mexico, South Korea, Taiwan, Thailand.
Recall number
Z-1401-2024
Recall status
Open, Classified
CDRH record ID
206264
FDA classified
2024-03-29
Recall event number
94083

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗