Skip to content

Public FDA record

Z-1402-2012: Non-DEHP Microbore Administration Set with Female Luer Adapter Packaged Assembly The Curlin Ambulatory

Recall record Z-1402-2012 for MOOG Medical Devices Group. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1402-2012
Company in FDA record
MOOG Medical Devices Group
FDA classification date
Apr 13, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Ambulatory administration sets used with the Curlin Ambulatory Pumps including the 4000 CMS Ambulatory Pump, 6000 CMS Ambulatory Pump and the Pain Smart IOD Ambulatory Pump may leak during use.

Source facts

Action
Moog Medical Device Group sent a Medical Device Recall Notification letter dated March 20, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to remove the affected product from their inventory, contact the distributor or Moog Customer Service at 1-800-970-2337, prompt #7, then return product to distributor or Moog. For additional questions regarding this recall call Moog's Customer Service Department at (800) 970-2337.
Initiated
2012-03-20
Root cause
Nonconforming Material/Component
Terminated
2012-07-26
Distribution
Worldwide Distribution - USA including AZ,,, CA,,, FL,,, GA,,, KS,,, IL,,, MA,,, MD,,, MI,,, OH,,, PA,,, TX,,, VA,,, WI,,, and the countries of Germany,,, Italy,,, Morocco,,, New Zealand,,, and the United Arab Emirates,,,
Recall number
Z-1402-2012
Recall status
Terminated
CDRH record ID
108265
FDA classified
2012-04-13
Recall event number
61446

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 21, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗