Open original FDA record ↗Recall record
- FDA identifier
- Z-1402-2017
- Company in FDA record
- Metrex Research, LLC.
- FDA classification date
- Mar 3, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Metrex is recalling the CaviWipes Extra Large because they may have been contaminated during the packaging process.
Source facts
- Action
- Metrex sent an Urgent Medical Device Recall letter dated January 20, 2017, to all affected customers to inform them that Metrex is recalling one lot of CaviWipes Extra Large because they may have been contaminated during the packaging process. Customers were informed that while handling sealed containers, a production worker sustained a minor finger cut which may have come in contact with the exterior surface of the container. Customers were instructed to contact Metrex Customer care at (800) 537-7123 to receive a RMA number. The RMA will allow for a quick return and replacement or credit. Customers were instructed to identify any customers that may have been shipped the affected product and contact their customers using the customer notification letter that Metrex issued. Metrex requests that those customers contact Metrex Customer Care at (800) 537-7123 and complete the attached acknowledgement form. For questions regarding this recall call 734-947-6700.
- Initiated
- 2017-01-20
- Root cause
- Packaging process control
- Terminated
- 2017-09-01
- Distribution
- Nationwide Distribution to PA, FL, IL, MN, TX, AL, NY
- Recall number
- Z-1402-2017
- Recall status
- Terminated
- CDRH record ID
- 152763
- FDA classified
- 2017-03-03
- Recall event number
- 76318