The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.
Source facts
Action
The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.
Initiated
2006-06-30
Root cause
Other
Terminated
2008-06-19
Distribution
Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Recall number
Z-1403-06
Recall status
Terminated
CDRH record ID
43931
FDA classified
2006-08-22
Recall event number
34487
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.