Skip to content

Public FDA record

Z-1403-06: Terumo Advanced Perfusion System 1 Roller Pump 6 inch diameter; Catalog No. 801041.

Recall record Z-1403-06 for Terumo Cardiovascular Systems Corp. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1403-06
Company in FDA record
Terumo Cardiovascular Systems Corp
FDA classification date
Aug 22, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The roller pump speed control knobs and the centrifugal pump speed control knobs may fail to function/change pump speed.

Source facts

Action
The firm replaced the graphics driver board on complaint units as user complaints were received. The firm notified all users of the problem and the presence of an alternative method of pump speed control via recall letter dated 6/30/06.
Initiated
2006-06-30
Root cause
Other
Terminated
2008-06-19
Distribution
Worldwide-USA, Australia, Belgium, Canada, China, Dominican Republic, Egypt, Germany, India, Indonesia, Japan, Malaysia, Mexico, Pakistan, Republic of China, Singapore, South Africa, South Korea, Thailand and United Arab Emirates.
Recall number
Z-1403-06
Recall status
Terminated
CDRH record ID
43931
FDA classified
2006-08-22
Recall event number
34487

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗