Z-1403-2015: Vitek 2 AST-P641 REF 418 590 Gram Positive Susceptibility Card, 20 cards per carton, IVD. Intended for

Recall record Z-1403-2015 for Biomerieux Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1403-2015
Company in FDA record
Biomerieux Inc
FDA classification date
Apr 8, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class III

Reason stated in the source

The outer carton label and package insert for the recalled product incorrectly lists Streptococcus pneumoniae as an organism for intended use.

Source facts

Action
Biomerieux, Incs sent an Urgent Product Correction Notice on February 25, 2015, to all affected customers. The letter identified the product, the problem, and the actions to be taken by the customer. Customers were instructed to do the following: " Please confirm this letter has been distributed and reviewed by all appropriate personnel within your organization. " For the referenced test kits (VITEK 2 ASTP635, ASTP640 or ASTP641), review the claimed organisms in the Package Insert antibiotic table to confirm which organisms can be tested. " Refrain from using the referenced test kits to determine the susceptibility of Streptococcus pneumoniae. " Reference your Laboratory Standard Operating Procedures to identify card type used for specific organisms. " Please store this letter with your bioMrieux instrument documentation. Please complete the attached Acknowledgement Form and return it to bioMrieux, Inc. Customers with questions were instructed to contact their local BioMerieux representative. For questions regarding this recall call 314-731-8526.
Initiated
2015-02-25
Root cause
Employee error
Terminated
2017-03-22
Distribution
Worldwide Distribution to United Kingdom and Turkey. NO U.S. distribution.
Recall number
Z-1403-2015
Recall status
Terminated
CDRH record ID
134238
FDA classified
2015-04-08
Recall event number
70645

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗