Open original FDA record ↗Recall record
- FDA identifier
- Z-1403-2018
- Company in FDA record
- Teleflex Medical Europe Ltd
- FDA classification date
- Apr 17, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
These one-way valves may disconnect at the joint between the two components that make up the device, which would cause an interruption of ventilation to the patient.
Source facts
- Action
- Teleflex initiated the recall by letter on 03/27/2018. The firm directed the consignee as follows: "1. If you have affected stock, immediately discontinue use and quarantine any products with the product code and lot numbers listed above. 2. To return product, complete the enclosed Recall Acknowledgement Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document the amount of product you have on hand for return. A customer service representative will contact you with a Return Goods Authorization (RGA) Number and will provide instructions for the return of products to Teleflex Medical. 3. If you have no affected stock, please complete the enclosed Recall Acknowledgment Form and fax it to 1-855-419-8507, Attn: Customer Service or email to recalls@teleflex.com. This will allow us to document your receipt of this letter."
- Initiated
- 2018-03-27
- Root cause
- Device Design
- Terminated
- 2021-05-04
- Distribution
- US Nationwide in the states of AU, CA, NZ
- Recall number
- Z-1403-2018
- Recall status
- Terminated
- CDRH record ID
- 162940
- FDA classified
- 2018-04-17
- Recall event number
- 79670