Skip to content

Public FDA record

Z-1403-2022: RayStation 9B SP1. For radiation treatment planning.

Recall record Z-1403-2022 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1403-2022
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA record creation date
Jul 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.

Source facts

Action
The firm contacted all customers via email on June 1, 2022. Customers were provided with a list of workflows where safety related information may not be propagated. Customers were instructed to be aware that treatment course information in RayCare may not always be up to date. Customers should verify critical information in RayStation or RayTreat before treatment course information in RayCare is used for critical decisions. Customers were asked to identify all installed units with the affected software version numbers. The issue will be resolved in mandatory software updates for RayStation, scheduled for market release in June 2022.
Initiated
2022-06-01
Root cause
Software change control
Terminated
2023-10-04
Distribution
Worldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
Recall number
Z-1403-2022
Recall status
Terminated
CDRH record ID
194595
FDA record created
2022-07-19
Recall event number
90498

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗