Five complaints received where surgeon was unable to remove and/or connect the inserter with the stem during surgery due to excess material remaining within the taper. Issue may lead to clinically significant extension of surgery to find a replacement.
Source facts
Action
On March 3, 2025, the firm notified affected customers via Urgent Medical Device Recall letters. Customers were informed of the excess metal on the inner taper of the stem due to a chipped manufacturing tool.
Customers were asked to notify all affected persons and assist their Zimmer Biomet sales representative in quarantining all affected product on hand. The products will be removed from your facility.
Initiated
2025-03-03
Root cause
Nonconforming Material/Component
Distribution
US distribution nationwide.
International distribution to Dominican Republic, Canada, India, Australia, Japan, EMEA, Korea, Taiwan, and Hong Kong.
Recall number
Z-1403-2025
Recall status
Open, Classified
CDRH record ID
212776
FDA classified
2025-03-20
Recall event number
96398