Open original FDA record ↗Recall record
- FDA identifier
- Z-1406-06
- Company in FDA record
- Radiometer America Inc
- FDA classification date
- Aug 26, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
QC Bias-a solution used in the manufacturing of the membrane had a different composition which introduces a bias of 20 mpH.
Source facts
- Action
- The firm contacted their customers and informed them of this problem and provided a workaround procedure, in a a letter dated 1/12/2005. A ''Field Action Notification'' (FAN) was issued to the firm''s Field Service personnel on 1/21/2005. The letter instructs their personnel to update the NPT7 application software on all units to V3.05 and to hand over a copy of the letter ''Changes related to the Qualicheck2+ and NPT7,'' to the user/customers. In addition, the firm''s personnel were instructed to remove any of the suspect cartridges found on-hand at each customer location; destroy them;and replace them with different lot numberd product.
- Initiated
- 2005-01-12
- Root cause
- Other
- Terminated
- 2011-09-14
- Distribution
- The recalled device component was distributed to a total of 56 customers, nationwide located in the following states: AL, AR, AZ, CA, CO, GA, ID, IL, KA, MA, ME, MI, MO, MS, NC, ND, NE, NH, NJ, NV, NY, OH, OK, PA, SC, SD, TX, UT & WA.
- Recall number
- Z-1406-06
- Recall status
- Terminated
- CDRH record ID
- 46580
- FDA classified
- 2006-08-26
- Recall event number
- 35664