Open original FDA record ↗Recall record
- FDA identifier
- Z-1406-2021
- Company in FDA record
- Centinel Spine, Inc.
- FDA record creation date
- Apr 9, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Centinel Spine learned through five customer complaints that the product was missing the "UP" etching on the superior plate of the prodisc C device. Per the Surgical Technique, the "UP" is one of the methods used to indicate the direction in which the surgeon should place the prodisc C into intradiscal space. There are also "UP" indications in the packaging and on instrumentation as additional guides to help the surgeon properly implant the device that were present for the affected product.
Source facts
- Action
- In March 2021, Centinel issued an Urgent Medical Device Recall notice to customers and distributers via letter and email notifying them the "UP" etching on the superior plate is missing. The "UP" is used to signify that the implant should be inserted into the intradiscal space with "UP" towards the head or in the cephalic position. In those devices with "UP" missing, there is a risk that the device could be placed upside-down within the intradiscal space.
- Initiated
- 2021-03-05
- Root cause
- Employee error
- Terminated
- 2021-08-02
- Distribution
- US Nationwide distribution in the states of CA, MD, DC, TN, GA, NC, PA, KY, NY, AZ, CO, OH, FL, MI.
- Recall number
- Z-1406-2021
- Recall status
- Terminated
- CDRH record ID
- 186233
- FDA record created
- 2021-04-09
- Recall event number
- 87530