Open original FDA record ↗Recall record
- FDA identifier
- Z-1407-2010
- Company in FDA record
- AliMed Corporation
- FDA classification date
- Apr 16, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Alimed became aware of issue via internal receiving inspection activities on 1/20/09. Firm states that subsequent to this observation, Customer Dissatisfaction inquiry #100192 was generated on 1/22/10. A single unit was received back by AliMed and 0 n/2/1/20 from a rehabilitation center that indicated that it had received a unit and it did not function upon its initial use. Failure of the alarm
Source facts
- Action
- Letter sent to customers stating that some units may have been subjected to excessive heat during the manufacturing process causing internal distortion to a key alarm component. The letter also request that the Health Care professional review records and notify all users to whom the product may have been issued and if any product has been issued outside of their facility, to forward this correspondence to them. Firm has enclosed a form for return of the product and issuance of a new product. For questions, please contact the firm using the information provided in the letter.
- Initiated
- 2010-02-12
- Root cause
- Device Design
- Terminated
- 2011-07-25
- Distribution
- All product was distributed domestically with the exception of one unit sold in Canada. AliMed sells and distributes this product to medical health professionals and distributors as well as directly to end users
- Recall number
- Z-1407-2010
- Recall status
- Terminated
- CDRH record ID
- 89165
- FDA classified
- 2010-04-16
- Recall event number
- 54783