Recall record Z-1407-2025 for FUJIFILM Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1407-2025
Company in FDA record
FUJIFILM Healthcare Americas Corporation
FDA classification date
Mar 20, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The incorrect computed patient age is showing in VX for patients less than 3 months old.

Source facts

Action
On February 11, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. ACTIONS TO BE TAKEN BY THE COMPANY Fujifilm will issue corrections for all impacted 7.4.x versions. The following patch releases will be made available: - 7.4.020 for 7.4.0x versions (immediately available) - 7.4.120 for 7.4.1x versions (available February 28, 2025) - 7.4.210 for 7.4.2x versions (available February 28,2025) ACTIONS THAT CAN BE TAKEN BY THE END-USER IMMEDIATELY Regarding this issue, Synapse PACS VX can safely be used by the end-user as long as; 1] the end-user is aware of the issue and, 2] takes appropriate actions when rendering a diagnosis for patients under 18 years of age. The following workarounds can be used: 1) The image acquisition date and date of birth are displayed on the Image. The end-user can use the dates to verify the age or, 2) Open the Series Viewer by double clicking on the thumbnail displayed on top the Synapse VX Image viewing window or the worklist. The Series Viewer Window will render the image on the server using the Server-side rendering (Zero Viewer component) which will calculate the age correctly. Alternatively, the user can switch to the Synapse PACS Zero Viewer/Thin Viewer (Server-side Rendering). The Synapse PACS Zero Viewer/Thin Viewer calculates the age correctly. ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1) Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 2) Please contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to schedule a patch update. Fujifilm regrets any inconvenience this action may cause and appreciates your understanding and cooperation. Please contact our field action coordinator, Jay Wigley, via email at hcusquality@fujifilm.com, if you have any further questions regarding this field action. Please contact Technical Assistance Center (TAC) at 1-888-385-4622 or 1-888-FU
Initiated
2025-02-11
Root cause
Software change control
Distribution
Domestic: AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MN, MO, MS, NC, NJ, NY, OH, OK, OR, TN, TX, VT, WA, WI, WV, WY. International: Australia, Belgium, Brazil, France, Germany, Portugal, UK
Recall number
Z-1407-2025
Recall status
Open, Classified
CDRH record ID
212627
FDA classified
2025-03-20
Recall event number
96360

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗