Skip to content

Public FDA record

Z-1408-2022: Nonsterile convenience kit: regard Item #830014016, LD00199P - C Section Mother.

Recall record Z-1408-2022 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1408-2022
Company in FDA record
ROi CPS LLC
FDA classification date
Jul 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits.

Source facts

Action
The recalling firm initiated the recall via telephone on 6/3/2022 to the direct consignee of Item 830019017, explained the reason for recall, and instructed the product be quarantined. The recalling firm also contacted the end user who had received distribution from the direct consignee. The end user was informed of the error and was requested to remove the affected lot from their inventory. The recalling firm worked with the end user to have the kits returned to the direct consignee where they were placed in quarantine. Regarding the recall of Item #830099008 and Item #830014016, the direct consignees were notified on 6/6/2022 via telephone requesting the affected kits be placed on hold and moved to a quarantined area. A follow-up email was sent to the recall coordinator for the direct consignees (which was the same recall coordinator) on 6/16/2022 which contained a correction notice. The email explained the issue and said the recalling firm has already performed rework on the kits in the inventory at their location (the two direct consignee locations) so no additional actions were required except for the completion of the reply form to indicate receipt of the notice. The attached Medical Device Correction Notice was dated 6/16/2022 and explained the issue, the affected product, the health risk assessment, and the actions taken by the recalling firm, which included rework of the products that had been returned to the direct consignees. The enclosed replay form was to be completed acknowledging the consignee has read and understands the notice.
Initiated
2022-06-03
Root cause
Process change control
Terminated
2023-04-14
Distribution
Distribution was made to MO. There was no foreign/military/government distribution.
Recall number
Z-1408-2022
Recall status
Terminated
CDRH record ID
194588
FDA classified
2022-07-19
Recall event number
90492

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗