Unexpected low cholesterol results-The Cholesterol calibrator (Lp(a)) component of the in vitro diagnostic kit for cholesterol testing, has provided patient results lower than the correct value.
Source facts
Action
The recalling firm notified consignees by phone and e-mail with an attached letter dated 07/06/06. The notification, flagged as ''Important Device Recall'' advised consignees to stop using the recalled product and that Lp(a) results showed a decrease in value of approximately 10% when the identified lots are used. Consignees were requested to return the enclosed response form for replacement product.
Initiated
2006-07-06
Root cause
Process control
Terminated
2006-09-14
Distribution
The product was sold to one distributor and direct medical accounts Nationwide.
Recall number
Z-1409-06
Recall status
Terminated
CDRH record ID
47089
FDA classified
2006-08-25
Recall event number
35885