Skip to content

Public FDA record

Z-1409-06: Lp(a) Calibrator, Cat No. 998-41491, 1 ml glass bottles, in cardboard box, 2 bottles/box ,Waco Chemicals

Recall record Z-1409-06 for Wako Chemicals, USA Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1409-06
Company in FDA record
Wako Chemicals, USA Inc
FDA classification date
Aug 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Unexpected low cholesterol results-The Cholesterol calibrator (Lp(a)) component of the in vitro diagnostic kit for cholesterol testing, has provided patient results lower than the correct value.

Source facts

Action
The recalling firm notified consignees by phone and e-mail with an attached letter dated 07/06/06. The notification, flagged as ''Important Device Recall'' advised consignees to stop using the recalled product and that Lp(a) results showed a decrease in value of approximately 10% when the identified lots are used. Consignees were requested to return the enclosed response form for replacement product.
Initiated
2006-07-06
Root cause
Process control
Terminated
2006-09-14
Distribution
The product was sold to one distributor and direct medical accounts Nationwide.
Recall number
Z-1409-06
Recall status
Terminated
CDRH record ID
47089
FDA classified
2006-08-25
Recall event number
35885

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗