Z-1410-06: INOvent, Nitric Oxide Delivery System. Model Number 1605-9000-000

Recall record Z-1410-06 for INO Therapeutics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1410-06
Company in FDA record
INO Therapeutics, Inc.
FDA classification date
Aug 25, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Kel-F Tip mounted on the high pressure hose can become dislodge/lodged in the INOmax valve outlet.

Source facts

Action
On June 29, 2006, the firm sent a letter to its consignees advising them to visually verify that the Kel F tip is in place before attempting to connect either of the high pressure hoses to the INOmax cylinders. INO advised that if the tip was not in place, the operator should immediately replace the tip before attaching the hoses. On July 5, 2005, the firm sent a second letter to their consignees with extra Kel F tips and existing instructions from the INOvent Operation & Maintenance Manual outlining the proper replacement procedure for the tips. The firm is following with telephone contact to assure that the letters were delivered to the proper person(s).
Initiated
2006-06-29
Root cause
Other
Terminated
2007-07-16
Distribution
Worldwide, USA and countries of Canada, Uruguay, Chile, Argentina, Colombia, and Malaysia
Recall number
Z-1410-06
Recall status
Terminated
CDRH record ID
47097
FDA classified
2006-08-25
Recall event number
35892

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗