External tubing may separate during use, allowing blood loss.
Source facts
Action
All of Merit's US sales representatives and distributors were notified by Prelude Short Sheath Introducer - Product Recall letter on 4/14/2009. They were instructed to promptly contact their customers. Customers were instructed to isolate, remove from inventory, and return all affected inventory. A Product Retrieval Form was to be returned to Merit. Further information is available at 801-316-4998 or 801-208-4344.
Initiated
2009-04-14
Root cause
Process control
Terminated
2010-02-09
Distribution
Worldwide Distribution -- USA, including states of CA, CO, GA, IL, MO, OK, and OR and countries of Austria, Belgium, Canada, Denmark, France, Germany, Hong Kong, Japan, Latvia, Lithuania, Netherlands, New Zealand, Spain, Sweden, and UK.
Recall number
Z-1410-2009
Recall status
Terminated
CDRH record ID
81520
FDA classified
2009-06-10
Recall event number
51877