Skip to content

Public FDA record

Z-1410-2021: Chesapeake Cervical-Ti 10mm Drill Guide Inserter- a multi-screw construct providing stability to the ante

Recall record Z-1410-2021 for K2M, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1410-2021
Company in FDA record
K2M, Inc
FDA record creation date
Apr 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Interference condition between the 10mm Drill Guide Inserter and the inner shaft that mates with the inserters. As a result, inner shafts do not mate flush with the collet inserters, and splaying of the collet. If implant falls into the spinal cord may cause a spinal cord injury or Incomplete seating of the inserter to the implant could result in misalignment of the screws with the inserter screw guide, potentially causing malposition of the screws into the adjacent vertebral bodies

Source facts

Action
Stryker Spine issued Urgent Medical Device Recal letter that issued on 3/16. Letter states reason for recall, health risk and action to take: Our records indicate Stryker previously distributed catalog no. 3608-90033, Lot MNAJ to your location in a Stryker Loaner Set. The nonconforming instrument was subsequently identified and quarantined during routine processing of loaner sets by Stryker at its Leesburg facility. This communication is being made for notification purposes only. The only action requested of you is to complete the Business Reply Form found on the following page and return it to the email address referenced on the form. If you have any questions or concerns, please contact Stryker Spine Regulatory Compliance at 201-749-8090. On behalf of Stryker we thank you sincerely for your help and support in completing this action by the target date March 19, 2021.
Initiated
2021-03-16
Root cause
Under Investigation by firm
Terminated
2021-08-12
Distribution
PA
Recall number
Z-1410-2021
Recall status
Terminated
CDRH record ID
186282
FDA record created
2021-04-12
Recall event number
87565

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗