Z-1410-2022: MODULIFT VBR SZ.M 4MM SET SCREW DRIVER, MF790R, part of Modulift VBR Medium Instrument Set, ST-0441

Recall record Z-1410-2022 for Aesculap Implant Systems LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1410-2022
Company in FDA record
Aesculap Implant Systems LLC
FDA record creation date
Jul 19, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Set screwdriver fails during surgery by deforming at the tip during engagement and manipulation of the implant.

Source facts

Action
An Urgent Medical Device Removal Notification was issued October 25, 2021 sent via email and USPS certified mail. The notice identified the product MODULIFT VBR SZ.M 4MM SET SCREWDRIVER, part of Modulift VBR Medium Instrument Set, ST-0441 and requested customers identify and quarantine affected inventory, submit the acknowledgement form and contact AIS Customer Service to arrange for return of product at (866) 229-3002 and SetWorld_orders.BBMUS_Service@aesculapimplants.com. Distributors are to notify customers if product was further distributed.
Initiated
2021-10-25
Root cause
Component design/selection
Terminated
2024-06-12
Distribution
US: CA, FL, GA, IL, OH & MI
Recall number
Z-1410-2022
Recall status
Terminated
CDRH record ID
194728
FDA record created
2022-07-19
Recall event number
90544

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗