Open original FDA record ↗Recall record
- FDA identifier
- Z-1411-2010
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Apr 22, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Firm became aware of a failure in the position detection system which could result in incorrect positioning of the detector in the longitudinal or lateral direction.
Source facts
- Action
- The following Important Notices were sent to customers: 1) Important Notice A305-Possible Synergy XVI panel position error (dated September 19, 2007) was sent to customers with versions R3.5, R4.0 and R4.1., 2) Important Notice A306-Possible Synergy XVI panel positioning error (dated November 8, 2007) was released to encompass customers with version R4.2., and 3) Important Notice A328 dated March 10, 2009, and sent again on April 29, 2009, included a modification kit to address the original issue and an expanded range of serial numbers affected. This Important notice A328 entitled, Elekta Synergy kV Detector Position Readback Modification Kit supersedes A305 and A306 for customers who have serial numbers within the range specified. The letter describes the product and the problem and advises the customer to follow any instructions or recommendations covered in this Notice. If you have any questions, please contact Elekta at 770-670-2548.
- Initiated
- 2007-09-19
- Root cause
- Device Design
- Terminated
- 2012-08-09
- Distribution
- Worldwide distribution: USA and Puerto Rico, Canada and Mexico.
- Recall number
- Z-1411-2010
- Recall status
- Terminated
- CDRH record ID
- 89892
- FDA classified
- 2010-04-22
- Recall event number
- 55099