Skip to content

Public FDA record

Z-1411-2021: Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessorie

Recall record Z-1411-2021 for Covidien. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1411-2021
Company in FDA record
Covidien
FDA record creation date
Apr 12, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

Source facts

Action
On 3/10/2021, Medtronic issued an Urgent Medical Device Recall notices to customers. All consignees were notified by letter delivered via Federal Express and/or certified mail. The letter informs customers of the recall of specific item codes and production lots of Covidien Surgiwand" II Suction and Irrigation Device following reports from customers of foreign particles in the device tubing.
Initiated
2021-03-05
Root cause
Process design
Distribution
Domestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador, Guatemala, Jamaica, Mexico, Panama, Peru, Puerto Rico, Trinidad And Tobago, Austria, Belgium, Bosnia And Herzegovina, Canary Islands, Cyprus, Denmark, Equatorial Guinea, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Kosovo, Kuwait, Lebanon, Mauritius, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Russian Federation, Rwanda, Saudi Arabia, South Africa, Spain, Swaziland, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom.
Recall number
Z-1411-2021
Recall status
Open, Classified
CDRH record ID
186254
FDA record created
2021-04-12
Recall event number
87542

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗